2 application-dependent settings, 1 general information, Application-dependent settings -2 4.2.1 – YSI IQ SensorNet TetraCon Sensor 700 IQ (SW) User Manual

Page 16: General information -2

Advertising
background image

Measuring / Operation

TetraCon

®

700 IQ (SW)

4 - 2

ba76018e01

01/2012

4.2

Application-dependent settings

4.2.1

General information

The TetraCon

®

700 IQ (SW) conductivity sensor is stable over the long

term. When being used for the authorized use of the sensor in water/
wastewater applications, it is immediately ready for use.

Note
Normally, the conductivity measuring cell does not age. Special
measuring mediums (e.g. strong acids and bases, organic solvents) or
temperatures that are too high may considerably reduce its lifetime or
lead to damage. No warranty claims can be made for mechanical
damage or any failure caused by these types of measuring mediums.

Adapting the cell

constant to the

installation location

In the case of special installation conditions, it may be necessary to
adapt the cell constant (due to the influence of the measuring
environment, e.g. of boundary fields).

Note
The cell constant is stored in the controller. It is automatically allocated
to the substitute sensor when the sensor is exchanged.

Cell constants with YSI

installation accessories

In chapter 7 T

ECHNICAL

DATA

of this operating manual you will find the

correct values to set for some products of the YSI accessory program
that require a correction of the cell constant (if it is possible to give fixed
values at all). If necessary, special installation recommendations for the
TetraCon

®

700 IQ (SW) can be found in the accessory operating

manual. In case of doubt it is necessary to determine the cell constant
according to section 4.2.2.

Note
The setting of the cell constant is made in the setting menu of the
sensor according to section 3.4.

Advertising