7

What to do if...

No measured value

Measurement does not

function

System cannot be

calibrated

Cause

Remedy

– Sensor not connected

– Connect the sensor

– Unknown

– Look in the log book

Cause

Remedy

– Watering cap still on the

electrode

– Pull off watering cap and

calibrate

– Electrode not connected

– Connect electrode

– Liquid has penetrated the

sensor

– Sensor defective, send it back

– Sensor not connected

– Connect the sensor

– Instrument setting incorrect

– Correct instrument setting

Cause

Remedy

– Slope of the electrode not

within tolerance (see section
4.2.5)

– Condition the electrode

– If the slope is still outside the

tolerance: replace electrode

– Slope of the electrode too low

– Replace electrode

– Asymmetry of the electrode

too high

– Replace electrode

– Sensor is operated with ORP

electrode

– Use pH electrode