4.9.2
Resetting the photometry settings
With the Photometry / Reset function, all photometry settings are reset.
4.9.3
Resetting the pH settings
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default
settings with the Reset function:
Setting
Default settings
PROG
0
Cell size
16 mm
Measured parameter
Concentration
Wavelength
436 nm
Blank values
all erased
Setting
Default settings
Cal. type
TEC
Calibration interval
7 d
Measured parameter
pH & ORP
Asymmetry (Asy)
0 mV
Slope (Slp.)
-59.16 mV/pH
Temperature, manual
25 °C
Temperature
°C