Configuring the RS232 interface

For an error-free data transmission, the RS232 interface should be set
to the same transmission speed (Baud rate) on the pHotoFlex

®

pH and

PC.

You can set the following values for the baud rate on the
pHotoFlex

®

pH : 1200, 2400, 4800, 9600, 19200.

The baud rate is selected in the menu, Configuration / System / Inter-
face
/ Baud rate.

4.8.3

Selecting the output format of datasets

For downloading data to the interface you can select the output format.

It is selected in the menu, Configuration / System / Interface / Output
format
.

The ASCII output format delivers formatted datasets.
The CSV output format delivers datasets separated by ";".

Output format, ASCII

pHotoFlex Ser. no. 12345678
31.10.04 09:56:20
ID: 1
10.01 (pH) AR
25 °C
________________________________________

pHotoFlex Ser. no. 12345678
31.10.04 15:48:08
ID 1 / PROG 2
1.1 mg/l Pb [BV][9 + 1]
________________________________________

etc...