YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 83

Advertising
background image

pHotoFlex

®

pH

What to do if...

83

ba76123e01

02/2013

Measured value display

----

(calibration error)

No stable measured

value

Cause

Remedy

Electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

– Moisture in the plug

– Dry plug

– Not enough electrolyte

– Top up electrolyte

– Electrode obsolete

– Replace electrode

– Electrode broken

– Replace electrode

– Socket damp

– Dry socket

Calibration procedure:

– Incorrect solution temperature

(without temperature sensor)

– Set up correct temperature

– Incorrect buffer solutions

– Select buffer solutions

suitable for the calibration
procedure

– Buffer solutions too old

– Use only once.

Note the shelf life

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Temper if necessary

Electrode + test sample:

– Conductivity too low

(e.g. in ultrapure water)

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode