YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 73

Advertising
background image

pHotoFlex

®

pH

Operation

73

ba76123e01

02/2013

Storing user-defined

programs

Enter a command line according to the following scheme in the terminal
program:

U.500#14,Program No,Model,Cell [mm],Wave length [nm],MRB,
MRE,E0,Slope,Resolution,Citation form,Unit,Blank required,Standard
adjustment possible
,Standard adjustment required

Example:

U.500#14,900,Test,16,436,0.0,2.0,0.0,1.0,0.01,test,

mg/l,0,1,0

The individual data sections of the command line are separated by
commas. The dot "." has to be used as a decimal separator within a
data section.

If a program is already stored under the selected number, the program-
ming procedure is canceled. To store the program under the selected
number, first erase the program stored under the number.

After successful transmission, the terminal program writes "!>".
If the transmission failed, the terminal program writes "!>".

Erasing user-defined

programs

To erase user-defined programs, enter a command line according to
the following scheme in the terminal program:

After successful transmission, the requested data is displayed on the
terminal.
If the transmission failed, the terminal program writes "!>".

5

Enter the command line.

6

Finish the command line with Enter.

The data is transmitted to the pHotoFlex

®

pH.

Erase all user-defined
programs

Erase one user-
defined program

Command line

U.520

U.521#1,Program No

Example

U.520

U.521#1,900

7

Enter the command line.

8

Finish the command line with Enter.

The data is transmitted to the pHotoFlex

®

pH.

The requested data is displayed on the terminal as the result.