YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 55

Advertising
background image

pHotoFlex

®

pH

Operation

55

ba76123e01

02/2013

Sample of a record

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.

31.10.03 16:13
pHotoFlex Ser. no. 12345678
Calibration pH & ORP
Calibration date

31.10.03 16:13:33

Calibration interval

7 d

AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope

-60.2 mV/pH

Asymmetry 4.0 mV
Sensor +++

Sample dis-
play

Calibration
record

Asymmetry
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the electrode according
to the electrode operating
manual

----

----

< -30 or
> 30

... -62 or
... -50

Eliminate the error according to
chapter 6 W

HAT