6 measuring the absorbance/transmission, Measuring the absorbance/transmission – YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 43

Advertising
background image

pHotoFlex

®

pH

Operation

43

ba76123e01

02/2013

4.5.6

Measuring the absorbance/transmission

The transmission measurement is not described separately in the fol-
lowing example as it operates in exactly the same way as the absor-
bance measurement. The result of the measurement is displayed in %
Transmission
.

1

Press the <M> key (long pressure) repeatedly until the Pho-
tometry
measuring mode is selected.

2

Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Absorbance or % Transmission is selected.

3

Select the cell diameter with <> <> and confirm with
<START/ENTER>.

4

Select the wavelength with <> <> and confirm with
<START/ENTER>.

5

Clean the cell (see section 5.2.2).

6

Insert the cell (see section 4.2).

7

Start the measurement with <START/ENTER>.

The measurement result is displayed when the measurement
is finished.

Photometry \ Absorbance

i Select cell
with

16 mm

610 nm

01.02.05 15:12

Photometry \ Absorbance
i Insert sample

i Start measurement
with <START>

16 mm

610 nm

01.02.05 15:12