3 ph value / orp voltage, Ph value / orp voltage – YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 88

Advertising
background image

Technical data

pHotoFlex

®

pH

88

ba76123e01

02/2013

7.3

pH value / ORP voltage

Measuring ranges,

resolution

Variable

Measuring range

Resolution

pH

- 2.00 ... + 16.00

0.01

U [mV]

- 1000 ... + 1000

1

T [°C]

- 5.0 ... + 100.0

0.1

T [°F]

- 23.0 ... + 212.0

0.1

Manual

temperature input

Variable

Range

Increment

T

manual

[°C]

- 20 ... + 100

1

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

pH *

± 0.01

+ 15 °C ... + 35 °C

U [mV]

± 1

+ 15 °C ... + 35 °C

T [°C]

± 0.3

0 °C ... + 55 °C

T [°F]

± 0.54

0 °C ... + 55 °C

* when measuring in a range of ± 2 pH around a calibration point