YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 42

Advertising
background image

Operation

pHotoFlex

®

pH

42

ba76123e01

02/2013

If a standard adjustment is stored for a program, the standard adjust-
ment is automatically used for measurement. The usage of the stan-
dard adjustment is documented together with the measured value and
indicated in the measured value display with [Cal].

6

Enter the nominal value of the standard.

Note:

Enter the decimal separator with <START/ENTER>.

7

Confirm the entered nominal value with <START/ENTER>.

8

Insert a cell with standard (see section 4.2).

9

Start the measurement of the standard with <START/ENTER>.

After measuring, the result of the standard adjustment is dis-
played and stored.
As the result, the measured value and the adjustment (in %) or
Error is displayed.

10

Confirm the result with <START/ENTER>.

The standard adjustment is completed.
The meter is ready to measure.

or:

Discard the result with <ESC>.
Subsequently, carry out a new standard adjustment.

mg/l Cu

_

Enter nominal value

i304: Cu-1 TP
Cell

i = 10 mm

i 0.600 mg/l Cu:

2.000 (93.2%)

Accept

Measure standard