YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 58

Advertising
background image

Operation

pHotoFlex

®

pH

58

ba76123e01

02/2013

Continuing for two-point

calibration

(Cal. type TEC)

8

Thoroughly rinse the electrode with distilled water.

9

Immerse the electrode in buffer solution 2.

10

If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).

11

Using <> <>, select Continue and press <START/EN-
TER>
. The buffer is measured.

The measured value is checked for stability (AutoRead).

12

Wait for the end of the AutoRead measurement or Terminate
AutoRead
with <START/ENTER> and take over the calibration
value.
The calibration display for the next buffer appears.

13

For two-point calibration, select Exit with 2 point calibration with
<> <> and confirm with <START/ENTER>.

The calibration is completed as a two-point calibration.

The new calibration values are displayed as an informative
message (

i

).

You have the following options:
 Accept the new calibration values with <START/ENTER>.

Subsequently, the calibration record is displayed and output
to the interface at the same time.

 To switch to the measured value display without accepting

the new calibration values, press <M> (short pressure) or
<ESC>.

i Buffer value = 10.01
i U = -177 mV
i Temperature =

24.8 °C

Terminate AutoRead

pH & ORP \ Calibration

i Buffer recognition TEC
i Immerse sensor in buffer 3

Exit with 2 point calibration

Continue

pH & ORP \ Calibration