Administrating user-defined methods

User-defined programs can be

 entered
 read out
 erased.

To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex

®

pH.

Administration of the method data can be carried out in the LSdata soft-
ware (see operating manual LSdata) or with the aid of a terminal pro-
gram (section 4.12.1).

You need the following data in any case:

Data

Possible entries/examples

Program No:

900 ... 999

Model:

Any name
(max. 5 characters)

Cuvette [mm]:

16 or 28

Wave length [nm]:

436, 517, 557, 594, 610 or 690

MRB:
(measuring range beginning)

e.g. 0.05

MRE:
(measuring range end)

e.g. 8.00

E0: (Offset)

e.g. 0.0

Slope: e.g.

1.0

Resolution:

0.0001, 0.001, 0.01, 0.1, 1, 2, 5, 10
or 100

Formula: (citation form)

e.g. PO4-P (max. 9 characters)

Unit

e.g. mg/l (max. 7 characters)

Blank required:
(No/Yes)

0 or 1

Standard adjustment possi-
ble
: (No/Yes)

0 or 1

Standard adjustment re-
quired
: (No/Yes)

0 or 1