Manganese vario, Mn-1 tp, 2 - 20.0 mg/l mn – YSI pHotoFlex Series User Manual

Page 38: 28 mm, Rn303, 00 manganese vario

Advertising
background image

38

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Mn-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.2 - 20.0 mg/l Mn

Display in mmol/l possible

Check the pH value of
the sample.
Desired range: pH 4-5.
Correct with diluted nitric
acid or sodium hydroxide
solution as necessary.

Pipette 10.0 ml of sample
into the empty cell.

Add the contents of a
VARIO MANGANESE
Citrate Buffer F10

powder pack and close
the cell with the screw
cap.

Mix the contents by
carefully swaying the cell
(10 x).

Add the contents of a
VARIO Sodium
Periodate F10
powder
pack and close the cell
with the screw cap.

Mix the contents by
carefully swaying the cell
(10 x).

Allow to react for 2
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

pH 4-5

2:00

Manganese vario

Program no.

303

*RN303*

Advertising