Chlorine (total) vario, Cl2-4 tp, 5 - 5.0 mg/l cl – YSI pHotoFlex Series User Manual

Page 26: 28 mm, Rn328

Advertising
background image

26

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Cl2-4 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.5 - 5.0 mg/l Cl

2

Display in mmol/l possible

Check the pH value of
the sample.
Required range: pH 6-7.
Correct with diluted
sodium hydroxide
solution or sulfuric acid
as necessary.

Pipette 10.0 ml of
sample into an empty
beaker.

Add the contents of a
VARIO Chlorine Total-
DPD F25 ml
powder
pack and dissolve them
by stirring.

Add 15.0 ml deionized
water with a pipette and
mix.

Fill an empty cell with the
prepared sample and
close it with the screw
cap.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

The test sample should be pink. Very high chlorine concentrations in the sample cause yellow solutions and too low mea-
sured values. Dilute the sample in this case.

Each time after determining total chlorine, rinse the cell with sulfuric acid 25 % and then several times with distilled water.

pH 6-7

Chlorine (total) vario

Program no.

328

*RN328*

Advertising