Chlorine (total) vario, Cl2-4 tp, 5 - 5.0 mg/l cl – YSI pHotoFlex Series User Manual

Page 26: 28 mm, Rn328

Advertising
background image

26

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Cl2-4 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.5 - 5.0 mg/l Cl

2

Display in mmol/l possible

Check the pH value of
the sample.
Required range: pH 6-7.
Correct with diluted
sodium hydroxide
solution or sulfuric acid
as necessary.

Pipette 10.0 ml of
sample into an empty
beaker.

Add the contents of a
VARIO Chlorine Total-
DPD F25 ml
powder
pack and dissolve them
by stirring.

Add 15.0 ml deionized
water with a pipette and
mix.

Fill an empty cell with the
prepared sample and
close it with the screw
cap.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

•

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

•

The test sample should be pink. Very high chlorine concentrations in the sample cause yellow solutions and too low mea-
sured values. Dilute the sample in this case.

•

Each time after determining total chlorine, rinse the cell with sulfuric acid 25 % and then several times with distilled water.

pH 6-7

Chlorine (total) vario

Program no.

328

*RN328*

Advertising