Copper vario, Cu-1 tp, 04 - 5.00 mg/l cu – YSI pHotoFlex Series User Manual
Page 33: 28 mm, Rn302, 00 copper vario
33
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:
Cu-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.04 - 5.00 mg/l Cu
Display in mmol/l possible
Check the pH value of
the sample.
Desired range: pH 4-6.
Correct with diluted
sodium hydroxide
solution or caustic
potash solution as
necessary.
Pipette 10.0 ml of sample
into the empty cell.
Add the contents of a
VARIO Cu1 F10 powder
pack and close the cell
with the screw cap.
Mix the contents by
carefully swaying the cell
(10 x).
Undissolved powder has
no adverse effect on
measurement.
Allow to react for 2
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
pH 4-6
2:00
Copper vario
Program no.
302
*RN302*