Ammonia (free), Mca-1 tp, 010 - 0.500 mg/l nh – YSI pHotoFlex Series User Manual

Page 16: N (f) 28 mm, Rn347

Advertising
background image

16

ba76126e01

03/2013

YSI model no.:

MCA-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.010 - 0.500 mg/l NH

3

-N (f)

0.012 - 0.610 mg/l NH

3

(f)

Display in mmol/l possible

Perform a zero
adjustment using a
28 mm cell filled with
sample.

Pipette 10.0 ml sample
into an empty 28 mm
cell.

Add 1 drop of Free
Ammonia Reagent

solution, close the cell
with the screw cap and
mix.

Allow to react for 5
minutes.

Add the contents of a
Vario Monochlor F
RGT
powder pack and
close the cell with the
screw cap.

Note: Failing to restore
the zero adjustment
will result in all future
measurements being
erroneous!

Shake the cell vigorously
for about 20 seconds to
dissolve solids.

Allow to react for 5
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

When switching to
another program restore
the zero adjustment by
performing a new zero
adjustment using a
28 mm cell filled with
deionized water.

Notes:

Collect samples in clean glass bottles and analyze as soon as possible.

The 5 minutes reaction times refer to a sample temperature of 18-20 °C (64-68 °F).
For other sample temperatures adjust the reaction time according to the follwing table:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

5:00

5:00

Sample temperature

Reaction time

(minutes)

Sample temperature

Reaction time

(minutes)

°C

°F

°C

°F

5

41

10

20

68

5

10

50

8

23

73

2.5

16

61

6

25

77

2

Ammonia (free)

Program no.

347

*RN347*