34

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

DEHA-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.004 - 0.450 mg/l DEHA

Display in mmol/l possible

Pipette 25.0 ml of
sample into an empty
beaker.

Add the contents of a
Vario Oxyscav 1 RGT
powder packet and
dissolve it by stirring.

Add 0.5 ml Vario DEHA
2 RGT
with a pipette and
mix.

Fill an empty cell with the
prepared sample, close it
with the screw cap and
put it in a dark place
.

Allow the sample to react
for ten minutes in a dark
place
. Then measure
immediately.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

Clean all laboratory glassware with hydrochloric acid (approx. 20 %), then thoroughly rinse with deionized water.

Avoid excessive movements and exposure to sun light during sampling. Store the samples hermetically sealed.

The temperature of the samples must be 25±3 °C.