Aluminium, Al-1 tp, 002 - 0.250 mg/l al – YSI pHotoFlex Series User Manual

Page 15: 28 mm, Rn323, 00 aluminium

Advertising
background image

15

ba76126e01

03/2013

Note: Before using this test with your photometer for the first time, determine the reagent blank
value (see notes below).

YSI model no.:

Al-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.002 - 0.250 mg/l Al

Pipette 20.0ml sample
into the empty cell.

Add the contents of a
VARIO Aluminum
ECR F20
powder pack
and close the cell with
the screw cap.

Dissolve the powder by
shaking.

Allow to react for 30
seconds.

Add the contents of a
VARIO Hexamine F20
powder pack and close
the cell with the screw
cap.

Dissolve the powder by
shaking.

For reagent blank value
only:
Add 2 drops of VARIO
ECR-Masking RGT
,
close the cell with the
screw cap and mix.

Allow to react for 5
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

To avoid errors due to contamination, rinse the equipment with hydrochloric acid solution (approx. 20%), then deionized
water beforehand.

Adding the

VARIO ECR-Masking RGT

masking reagent is only required when determining the reagent blank value.

The sample temperature has to be between 20 and 25 °C.

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

5:00

Aluminium

Program no.

323

*RN323*