Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Ntot2 TC (HR)

Category:

KT (reaction cell test)

Cell:

16 mm

Measuring range:

10 - 150 mg/l N

Add the contents of a
VARIO TN Persulfate
Rgt.
powder pack into a
TN Hydroxide HR
digestion cell.

With a pipette add 0.5 ml
of sample, close the cell
with the screw cap and
mix vigorously for at least
30 s. A small amount of
solids may remain
undissolved.

Heat the cell in the
thermoreactor at 120 °C
for 30 minutes.

Remove the cell from the
thermoreactor and let it
cool down in a cell rack.

Add the contents of a
VARIO TN Reagent A
powder pack, close the
cell with the screw cap
and mix for at least 15 s.

Allow to react for 3
minutes.

Add the contents of a
VARIO TN Reagent B
powder pack, close the
cell with the screw cap
and mix for at least 15 s.

Allow to react for 2
minutes.

Pipette 2.0 ml of
prepared sample into a
TN Acid LR/HR
(Reagent C)
reaction
cell and close the cell
with the screw cap.

Mix the contents by
carefully swaying the cell
(10 x / for approx. 30 s
altogether).
Caution, the cell
becomes warm!

Note:
Always clean the powder funnel prior to adding a
reagent!

Allow to react for 5
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.