Phosphate, ortho, Po4-2 tc, 06 - 5.00 mg/l po – YSI pHotoFlex Series User Manual

Page 54: 16 mm, Rn315, 00 phosphate, ortho

Advertising
background image

54

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

PO4-2 TC

Category:

KT (reaction cell test)

Cell:

16 mm

Measuring range:

0.06 - 5.00 mg/l PO

4

0.02 - 1.63 mg/l PO

4

-P

Display in mmol/l possible

Pipette 5.0 ml of sample
into a reaction cell and
close the cell with the
screw cap.

Mix the contents by
carefully swaying the
cell.

Add the contents of a
VARIO Phos 3 F10
powder pack and close
the cell with the screw
cap.

To dissolve solids, shake
the cell for approx. 10 to
15 seconds. A small
amount of solids may
remain undissolved.

Allow to react for 2
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

2:00

Phosphate, ortho

Program no.

315

*RN315*

Advertising