Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Mn-2 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.007 - 0.700 mg/l Mn

Display in mmol/l possible

Check the pH value of
the sample.
Required range: pH 4-5.
Correct with diluted nitric
acid or sodium hydroxide
solution as necessary.

Pipette 10.0 ml of sample
into the empty cell.

Add the contents of a
Vario Ascorbic Acid
powder packet and close
the cell with the screw
cap.

Mix the contents by
carefully swaying the
cell.

Add 15 drops of Vario
Alkaline-Cyanide
Reagent Solution
and
close the cell with the
screw cap.

Mix the contents by
carefully swaying the
cell.

Add 21 drops of Vario
PAN Indicator Solution
0.1%
and close the cell
with the screw cap.

Mix the contents by
carefully swaying the
cell.

Allow to react for 2
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

Clean all laboratory glassware with nitric acid, then thoroughly rinse with deionized water.