Molybdenum vario, Mo-2 tp, 3 - 40.0 mg/l mo – YSI pHotoFlex Series User Manual

Page 41: 28 mm, Rn331, 00 molybdenum vario

Advertising
background image

41

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Mo-2 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.3 - 40.0 mg/l Mo

Display in mmol/l possible

Check the pH value of
the sample.
Nominal value: approx.
pH 7.
Correct with diluted
sodium hydroxide
solution or nitric acid as
necessary.

Pipette 25.0 ml of
sample into an empty
beaker.

Add the contents of a
Vario Molybdenum
HR 1 F25 ml
powder
pack and dissolve them
by stirring.

Add the contents of a
Vario Molybdenum
HR 2 F25 ml
powder
pack and dissolve them
by stirring.

Add the contents of a
Vario Molybdenum
HR 3 F25 ml
powder
pack and dissolve them
by stirring.

Allow to react for 5
minutes.

Fill an empty cell with the
prepared sample and
close it with the screw
cap.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

If any molybdenum is present, the solution develops a yellow color after all reagents have been added.

approx. pH 7

5:00

Molybdenum vario

Program no.

331

*RN331*

Advertising