Nitrate, No3-1 tc, 2 - 30.0 mg/l no – YSI pHotoFlex Series User Manual

Page 44: 16 mm, Rn314, 00 nitrate

Advertising
background image

44

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

NO3-1 TC

Category:

KT (reaction cell test)

Cell:

16 mm

Measuring range:

0.2 - 30.0 mg/l NO

3

-N

1.0 - 133.0 mg/l NO

3

Display in mmol/l possible

Pipette 1.0 ml of sample
into a reaction cell and
close the cell with the
screw cap.

Mix the contents by
carefully swaying the cell
(10 x).

Add the contents of a
VARIO Nitrate
Chromotropic
powder
pack and close the cell
with the screw cap.

Mix the contents by
carefully swaying the cell
(10 x). A small amount of
solids may remain
undissolved.

Allow to react for 5
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

5:00

Nitrate

Program no.

314

*RN314*