Nitrite vario, No2-1 tp, 002 - 0.300 mg/l no – YSI pHotoFlex Series User Manual

Page 47: 28 mm, Rn305, 00 nitrite vario

Advertising
background image

47

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

NO2-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.002 - 0.300 mg/l NO

2

-N

0.007 - 0.985 mg/l NO

2

Display in mmol/l possible

Pipette 10.0 ml of sample
into the empty cell.

Add the contents of a
VARIO Nitri 3 F10
powder pack and close
the cell with the screw
cap.

Shake the cell.
Undissolved powder has
no adverse effect on
measurement.

Allow to react for 15
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

15:00

Nitrite vario

Program no.

305

*RN305*