Silica lr vario, Si-1 tp (lr), 01 - 1.60 mg/l sio – YSI pHotoFlex Series User Manual

Page 60: 28 mm, Rn321

Advertising
background image

60

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Si-1 TP (LR)

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.01 - 1.60 mg/l SiO

2

0.005 - 0.748 mg/l Si

Display in mmol/l possible

Pipette 10.0 ml of sample
into the empty cell.

Add 15 drops of Vario
Molybdate 3 Reagent
Solution
and close the
cell with the screw cap.

Mix the contents by
carefully swaying the
cell.

Allow to react for 4
minutes (temperature
dependence, see note).

Add the contents of a
VARIO Silica Citric
Acid F10
powder pack
and close the cell with
the screw cap.

Mix the contents by
carefully swaying the
cell.

Allow to react for 1
minute (temperature
dependence, see note).

Add the contents of a
VARIO LR Silica Amino
Acid F F10
powder
pack, close the cell with
the screw cap and mix.

Allow to react for 2
minutes. If SiO

2

is

present in the sample,
the solution becomes
blue.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

The above mentioned reaction times are valid at room temperature (20 °C). At 10 °C the reaction time has to be doubled,
at 30 °C it has to be halved.

1:00

2:00

Silica LR vario

Program no.

321

*RN321*

Advertising