Wiring schematic #1, Electromagnetic flow meter – Xylem Magflux Electromagnetic Flow Meter User Manual
Page 35

35
Electromagnetic Flow Meter
FW: 842018/842009
Wiring Schematic #1
for "Remote Converter with Connection Box on Flow Sensor" on the op-
posite page (page
).
Flow Sensor
Connect one end of
the MJK cable to the
flow sensor as de-
scribed in the table
below. All 3 shields are
twisted and connected
in slot 1. The white
lead in pair no. 3 is not
used.
B Connection for remote converter
Terminal Signal
Wire
Colour before
March 2001
Colour from
March 2011
1
GND
Shield
-
-
2
E1
Pair no. 1
Red
Red
3
E2
Pair no. 1
White
White
4
GND
Pair no. 3
Black
Black
-
-
Pair no. 3
White
-
5
L1
Pair no. 2
Green
Blue
6
L2
Pair no. 2
White
Brown
7, 8, 9
not used
Remote Converter
(and Display Unit)
Connect the other end
of the MJK cable to
the remote converter
as shown to the right.
The "1 Shield" lead
goes into "Ground
from flow pipe", the
"2 Red" lead goes
into "E1 Red", the "3
White" lead into "E2
White", and so on.