Xylem Magflux Electromagnetic Flow Meter User Manual
Page 46

46
EN 3.05 MagFlux Manual 1212
Electromagnetic Flow Meter
1. Apply power to the sensor.
2. When the "Sensor Calibration Code" menu appears (see below), enter
the calibration code read from the converter label and press OK.
Sensor Calibration Code
|
Back
Select
>
S/N not entered
Set Flow Direction
Follow Arrow A
Follow Arrow B
Back
Select
OK
MagFlux1
Flow 0.00 m
3
/h
FTot 00.000 m
3
Select Setup
Service Menu
Sensor Calibration Code
Converter SW Ver.
Product Data
Reset Counter Time
Internal Meas. & Cal.
Minimum Velocity
Calibrate Flow now
Calibrate Volume
Calibrate mA
Test Velocity
Freeze Coil
Read Event Log
Back on Stock
Back
Select
OK
Set Flow Direction
Follow Arrow A
Follow Arrow B
Back
Select
OK
Sensor Calibration Code
Sensor Calibration Code
dw5uq4
Back
Select
OK
Sensor Calibration Code
A
B
A
B
3. Select the correct flow direction from the "Set Flow Direction" menu
and press OK.
4. Sensor calibration code configuration is now finished, and you are
returned to the Main Menu.