Electromagnetic Flow Meter

FW: 842018/842009

the Display Unit to each converter one by one. If neglected, a conflict be-
tween the units will arise when the RS 485 serial loop is established.
Change for example "Device 1" to "Device 2" as follows:
1. Press OK and use the up/down keys to change "1" to "2".
2. Press OK.
The MagFlux

®

converter ID is now changed to "2", and more converters

can be added to the Display Unit with the "Add Device" option.

Remove Device
If a device is to be removed from the Display Unit, proceed as follows:
1. Select the required device with the up/down keys.
2. Click OK to remove it.
The device is now removed from the Display unit and display returns to the
"Modbus" menu.

Time Format

24 Hour

AM/PM

Back Select

OK

Change Device Address

Device 1

Back Select

OK

Add Device

Change Device Address

Remove Device

Change Display ID No.

Set No. of Retries

Back Select

OK

Scanning

Back

Display 841xxx-xxx

Build Sep 15 2006 11:18:24

Language 2006 09 21

Back

Set Clock

Language

Modbus

Display SW Version

Add Device

Devices Found

Device 1

Back

OK

Change Device Addresse

Change Device 1

1

Back Select

OK

Remove Device

Device 1

Back Select

OK

Remove Device

Language

Set Clock

Modbus

Factory Settings

Display SW Version

Back Select

OK

Set Clock

2006-09-19 12:43:11

Back Select

>

Recover Factory Settings

Cancel

Delete Log

Restore Default Language

Restore Factory Settings

Back Select

OK

Factory Settings

Display Setup

Password

Specify Main Screen

Factory Settings

Data Logger

Set Sensor Name

Display Setup

Converter Setup

Main Menu

.

.

Service Menu

.

Language

(GB) English

(US) English

Dansk

Deutsch

Español

Français

Italiano

Nederlands

Norsk

Soumalainen

Svenska

Back Select

OK

Change Display ID No.

1

Back Select

OK

Change Display ID No.

Number of Displays

1

Back Select

OK

Set No. of Retries

005

This will reset the display.

Back Select

>

Set No. of Retries

Change Display ID No.
The display ID number and the number of displays may have to be
changed to avoid conflicts on the Modbus.

Set No. of Retries
If the data communication lines are subject to noise or other disturbances,
the number of retries can be raised to increase the chances of a succesful
change of settings.