16

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

Dissolved Oxygen Transmitter

Oxix

®

Sensor

Dissolved oxygen saturated

0 - 30.0 mg/l or ppm (parts per million)

0 - 120 %

Measurement principle

Optical, near infra-red fluorescence (l

in

= 475 nm, I

out

= 600 nm)

Sensor check

Automatic self-diagnostics

Pressure

6 bar (100 psi)

Materials

Boby and Head: PU, PVC, silicone, epoxy and stainless steel 316 SS

Optical lens: Sapphire

O-ring: Viton

Cable: PUR

Cleaning system

Built-in air or water jet

Cable

4 x 0,326 mm

2

, shielded, outer diameter Ø 5.0 mm (22 AWG)

Cable length

10 m (33 ft.), can be extended up to 600 mm

Response time

t

90

less than 1 second, t

95

less than 60 second

Flow velocity

No flow required

Power supply

12 V DC (10 -16 V DC), approx. 10 mA

Output

RS 485, 9600 baud, 2-wire

Accuracy

Better than 1% of actual reading (0,02 mg/l)

Resolution

Better than 0,02 mg/l

Temperature range

0 - 60 ºC

Temperature compensation

0 - 50 ºC

Enclosure rating

IP 68 to IEC 529 (10 m), NEMA 6

Dimensions

50 x 130 mm (diameter x length)

Mounting

1,5 inch NPT thread

0,25 inch NPT thread for jet spray cleaning tube

Weight

0,77 kg (1.7 lbs)