Xylem Oxix Dissolved Oxygen Transmitter User Manual

Page 26

Advertising
background image

26

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

Dissolved Oxygen Transmitter

7. See on the following picture where two jumpers must be inserted for a 115

V AC configuration:

115 V AC dual jumper positions

8. Insert and solder 2 jumpers (wires) in the positions indicated by the black

arrows.

9. Turn around the PCB and replace the 63 mAT fuse on the right side with a

125 mAT fuse ( or vice versa going from 115 VAC to 230 VAC).

10. Re-insert the PCB and re-assemble the converter (see also steps 4, 3, 2

and 1).