Xylem Oxix Dissolved Oxygen Transmitter User Manual
Page 26
Advertising
26
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
7. See on the following picture where two jumpers must be inserted for a 115
V AC configuration:
115 V AC dual jumper positions
8. Insert and solder 2 jumpers (wires) in the positions indicated by the black
arrows.
9. Turn around the PCB and replace the 63 mAT fuse on the right side with a
125 mAT fuse ( or vice versa going from 115 VAC to 230 VAC).
10. Re-insert the PCB and re-assemble the converter (see also steps 4, 3, 2
and 1).
Advertising