Physical sensor installation, Preparation, Positioning the sensor – Xylem Oxix Dissolved Oxygen Transmitter User Manual

Page 20: Preparation positioning the sensor

Advertising
background image

20

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

Dissolved Oxygen Transmitter

Physical Sensor Installation

Preparation

Select a position that can produce a valid and representative measuring value
and provide sufficient space for physical cleaning of the sensor on a regular
basis.

Positioning the Sensor

The positioning of the sensor is very application dependant. There are, how-
ever, some "do"s and "don't"s that can be used as guidelines.

Do

In

aeration tanks or basins with significant fluctuation in depth the

sensor should be mounted on a foating device that keeps the sensor
submerged to a fixed depth from the surface.
In tanks or basins with no or little fluctuation in depth the sensor may

be mounted in a fixed position approximately 1/4 below the surface.
The optimal position at installations in flow-through aeration basins with

one measuring point is near the tank inlet.
The optimal positions at installations in flow-through aeration basins

with two measuring points are near the tank inlet and the tank outlet.

Don't

Do not position the sensor at or very close to the surface. Strong wind

aeration leads to increased dissolved oxygen concentration in the sur-
face layer which may not be representative.
Do not use a fixed sensor position in aeration tanks/basins with signifi-

cant depth fluctuations. The dissolved oxygen saturation varies with
the depth.

Advertising