Physical sensor installation, Preparation, Positioning the sensor – Xylem Oxix Dissolved Oxygen Transmitter User Manual

Page 20: Preparation positioning the sensor

Advertising
background image

20

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

Dissolved Oxygen Transmitter

Physical Sensor Installation

Preparation

Select a position that can produce a valid and representative measuring value
and provide sufficient space for physical cleaning of the sensor on a regular
basis.

Positioning the Sensor

The positioning of the sensor is very application dependant. There are, how-
ever, some "do"s and "don't"s that can be used as guidelines.

Do

In

•

aeration tanks or basins with significant fluctuation in depth the

sensor should be mounted on a foating device that keeps the sensor
submerged to a fixed depth from the surface.
In tanks or basins with no or little fluctuation in depth the sensor may

•

be mounted in a fixed position approximately 1/4 below the surface.
The optimal position at installations in flow-through aeration basins with

•

one measuring point is near the tank inlet.
The optimal positions at installations in flow-through aeration basins

•

with two measuring points are near the tank inlet and the tank outlet.

Don't

Do not position the sensor at or very close to the surface. Strong wind

•

aeration leads to increased dissolved oxygen concentration in the sur-
face layer which may not be representative.
Do not use a fixed sensor position in aeration tanks/basins with signifi-

•

cant depth fluctuations. The dissolved oxygen saturation varies with
the depth.

Advertising