Xylem Oxix Dissolved Oxygen Transmitter User Manual

Page 55

Advertising
background image

55

Dissolved Oxygen Transmitter

GB Oxix Manual 080915

FW: 841013-009, 846001-002, 841513-009

A typical cleaning interval for aeration basins would be 2 hours (120 minutes).
In colder climates, condensation may form and then freeze in the tubing. This
can be prevented by setting the interval to 5 - 20 minutes.
The cleaning time would typically be 30 seconds.
The measurement delay (recovery time) may also be used to coordinate with
other processes. The delay merely freezes the last reading before cleaning
begins. A typical value would be 60 seconds.

Averaging

Units

mA output

Converter setup

20 mA output

20 mA =

25.00 mg/l

Back Select

OK

D.O. Units

%

mg/l

ppm

Back Select

OK

Averaging

Units

mA output

Cleaning

High alarm

Low alarm

Sensor error

Status

Factory settings

Service menu

Back Select

OK

4mA output

4 mA =

00.00 mg/l

Back Select

OK

When ”n/a” is displayed, the

menu is not available or not

applicable

Yes

Enable mA output?

Yes

No

Back Select

OK

Averaging

0001 sec

(from 1 sec to 120 min)

Back Select

>

Temperature unit

°C

°F

Back Select

OK

Temperature

This menu is for display only.

Clean now

Yes

No

Back Select

OK

Reset Maintenance

Yes

No

Back Select

OK

Maintenance

Runtime 1000 hour

Count

500

Back Select

OK

Clean now

Select cleaning interval

060 Min

(20 - 255 Min)

Back Select

>

Cleaning

Interval

Use DI to clean

Clean now

Maintenance

Back Select

OK

Interval

Off

On

Back Select

OK

Use DI to clean

Interval

Maintenance

DI function

NO

NC

Back Select

OK

Use DI to clean

No

Yes

Back Select

OK

Delay

00100 ms

Back Select

>

Electro-mechanical relay

Opto relay

Not in use

Back Select

OK

Cleaning time

005 sec

Back Select

OK

Cleaning time

005 sec

Back Select

OK

Measurement delay

10 sec

Back Select

OK

GB Oxix menuer #1 Converter setup 1

FW: 841013-009, 846001-002, 841513-009

Page 2 of 7 15. september 2008