Dissolved Oxygen Transmitter

Sensor

Error - General sensor error that may also
be transferred to a relay output. Contact
MJK.

Sensor communication

Error - Modbus communication errrors nor-
mally imply that a device has been moved,
added, removed or has changed status in an
illegal manner. Rearranged cables or wires
can also invoke this error message.
Reset the system to cancel the error condi-
tion.

Sensor hardware

Error - Change the sensor hardware.

Sensor setting

Error - Some settings are wrong or have
been changed.

Sensor supply

Error - The sensor supply is too low or too
high. Check the sensor cable connection.

System

Error - Internal VCC error. See the event log
for details.

System error

Error - The system software is inoperable in-
cluding faulty alarm and sensor bits. Contact
MJK.

D.O. offset

Error - A zero-point sensor problem. The
sensor needs a zero-point calibration.
Contact MJK.

Warnings

Sensor range exceeded

Warning - The measured value is out of
range. Check the measuring range and the
measuring units.

Info

Measurement hold

Info - The measurement is set on hold dur-
ing the cleaning cycle or by an error

.

Sensor needs cleaning

Info - The sensor needs cleaning.

System reset occurred

Info - The converter has been subject to a
software or hardware reset.