Xylem MULTI 3400i pH/Cond MULTI-PARAMETER METERS User Manual

Page 17

Advertising
background image

pH/Cond 3400i

Operation

17

Temperature measure-

ment in pH measure-

ments

You can perform pH measurements with or without a temperature sen-
sor as well as with the temperature sensor of a conductivity measuring
cell. The measuring instrument recognizes which sensors are connect-
ed and automatically switches to the correct mode for the temperature
measurement.

The following cases are distinguishable.

* If you do not wish that, you can:

– either disconnect the 2nd sensor and use the manual temperature

input or

– use an electrode with a temperature sensor.

If a temperature sensor is connected, it is indicated on the display by
TP.

Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solu-
tion manually using the <

> <> keys while keeping the <RUN/EN-

TER> key depressed.

Temperature sensor

Display

Mode

pH

Cond

yes

-

TP

Automatic with
pH temperature sensor

yes

yes

TP

-

-

Manual

-

yes

TP
flashes

The temperature value of the
2nd sensor (Cond) in the
same test sample is used for
the pH measurement*