Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature sen-
sor as well as with the temperature sensor of a conductivity measuring
cell. The measuring instrument recognizes which sensors are connect-
ed and automatically switches to the correct mode for the temperature
measurement.
The following cases are distinguishable.
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
If a temperature sensor is connected, it is indicated on the display by
TP.
Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solu-
tion manually using the <
▲> <▼> keys while keeping the <RUN/EN-
TER> key depressed.
Temperature sensor
Display
Mode
pH
Cond
yes
-
TP
Automatic with
pH temperature sensor
yes
yes
TP
-
-
Manual
-
yes
TP
flashes
The temperature value of the
2nd sensor (Cond) in the
same test sample is used for
the pH measurement*