6 setting the temperature compensation, Setting the temperature compensation – Xylem MULTI 3400i pH/Cond MULTI-PARAMETER METERS User Manual

Page 40

Advertising
background image

Operation

pH/Cond 3400i

40

4.4.6

Setting the temperature compensation

The calculation of the temperature compensation is based on the pre-
set reference temperature, Tref 20 or Tref 25 (see
section 4.8 C

ONFIGURATION

).

You can select one of the following temperature compensations:

z Nonlinear temperature compensation (nLF)

according to EN 27 888

z Linear temperature compensation (Lin) with adjustable coeffi-

cient in the range 0.001 ... 3.000 %

/

K

z No temperature compensation (----)

Note

ips

Select the following temperature compensations given in the table ac-
cording to the respective test sample:

Application tips

Test sample

Temperature compensation

Display in-
dicator

Natural water (ground
water, surface water
and drinking water)

nLF
according to DIN 38404
EN 27 888

Ultrapure water

nLF
according to DIN 38404
EN 27 888

Other aqueous solu-
tions

Set linear temperature coeffi-
cient
0.001 ... 3.000 %/K

Salinity
(seawater)

Automatically nLF according
to IOT

Sal,

Advertising