4.4.6
Setting the temperature compensation
The calculation of the temperature compensation is based on the pre-
set reference temperature, Tref 20 or Tref 25 (see
section 4.8 C
You can select one of the following temperature compensations:
z Nonlinear temperature compensation (nLF)
according to EN 27 888
z Linear temperature compensation (Lin) with adjustable coeffi-
cient in the range 0.001 ... 3.000 %
/
K
z No temperature compensation (----)
Note
ips
Select the following temperature compensations given in the table ac-
cording to the respective test sample:
Application tips
Test sample
Temperature compensation
Display in-
dicator
Natural water (ground
water, surface water
and drinking water)
nLF
according to DIN 38404
EN 27 888
Ultrapure water
nLF
according to DIN 38404
EN 27 888
Other aqueous solu-
tions
Set linear temperature coeffi-
cient
0.001 ... 3.000 %/K
Salinity
(seawater)
Automatically nLF according
to IOT
Sal,