4 conductivity/salinity, 1 general information, Conductivity/salinity 4.4.1 – Xylem MULTI 3400i pH/Cond MULTI-PARAMETER METERS User Manual

Page 33: General information

Advertising
background image

pH/Cond 3400i

Operation

33

4.4

Conductivity/Salinity

4.4.1

General information

Note
Conductivity measurements with the pH/Cond 340 can only be carried
out using the TetraCon 325 measuring cell.

Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.

Preparatory activities

Perform the following preparatory activities when you want to measure:

Temperature sensor

The TetraCon 325 conductivity measuring cell has a temperature sen-
sor integrated in it. The temperature sensor is shown on the display by
TP.

Temperature

compensation

Selectable linear or nonlinear temperature compensation, can be
switched off (see section 4.8 C

ONFIGURATION

).

Reference temperature,

Tref

The reference temperature (Tref) can be switched between 20 °C and
25 °C. It appears on the display as Tref20 or Tref25. To switch over the
reference temperature, see S

WITCHING

OVER

THE

REFERENCE

TEMPERA

-

TURE

, page 60.

1

Connect a conductivity measuring cell to the measuring instru-
ment. The instrument automatically switches over to conductiv-
ity measurement (status display

or SAL).

If the conductivity measuring cell is already connected, press
the <M> key repeatedly until the status display

or SAL ap-

pears.

2

Check the selected cell constant (see below) or calibrate the
measuring instrument with the measuring cell (see
section 4.4.5).

3

Using <

> <>, toggle between the measuring modes, con-

ductivity (

in

µS/cm) or salinity (SAL).