2 conductivity system messages, 3 general errors, Conductivity system messages – Xylem MULTI 3400i pH/Cond MULTI-PARAMETER METERS User Manual
Page 69: General errors

pH/Cond 3400i
What to do if...
69
6.2
Conductivity system messages
Error message
OFL
OFL
OFL
OFL
Error message
E3
E3
E3
E3
6.3
General errors
Display LoBat
Instrument does not
react
to keystroke
Display
to
to
to
to
Probe symbol flashes
Cause
Remedy
The measured value lies outside
the measuring range
– Measuring cell not connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Unsuitable calibration
solution
– Check calibration solutions
Cause
Remedy
– Batteries almost empty
– Exchange the batteries (see
section 5.1 M
AINTENANCE
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Switch the instrument on
while pressing the <CAL> key
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system