Dynamic accuracy, Stability – Atec Agilent-E5063A User Manual
Page 14
14
Dynamic accuracy
1
Description
Speciication
Typical
Magnitude
6 dBm
–30 dBm
–100 dBm
–110 dBm
± 0.31 dB
± 0.056 dB
± 3.83 dB
± 5.00 dB
Phase
6 dBm
–30 dBm
–100 dBm
± 11.8 deg
± 0.37 deg
± 33.6 deg
1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level.
Magnitude
Phase
Stability
1
Description
Speciication
Typical
Magnitude
Transmission:
100 kHz to 300 kHz
300 kHz to 6 GHz
6 to 12 GHz
12 to 18 GHz
Relection:
100 kHz to 300 kHz
300 kHz to 6 GHz
6 to 12 GHz
12 to 18 GHz
± 0.02 dB/°C
± 0.01 dB/°C
± 0.025 dB/°C
± 0.04 dB/°C
± 0.02 dB/°C
± 0.02 dB/°C
± 0.035 dB/°C
± 0.05 dB/°C
Phase
Transmission:
100 kHz to 300 kHz
300 kHz to 6 GHz
6 to 12 GHz
12 to 18 GHz
Relection:
100 kHz to 300 kHz
300 kHz to 6 GHz
6 to 12 GHz
12 to 18 GHz
± 0.4 deg/°C
± 0.2 deg/°C
± 0.5 deg/°C
± 0.6 deg/°C
± 0.4 deg/°C
± 0.2 deg/°C
± 0.5 deg/°C
± 0.6 deg/°C
1. Stability is defined as a ratio measurement at the test port.