7

Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration (ECal) module

Network analyzer : E5063A
Calibration kit

: 85093C (3.5 mm, 50 Ω) Electronic calibration (ECal) module

Calibration

: Full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration is not performed

Specification (dB)

Description

300 kHz to 10 MHz

10 MHz to 3 GHz

3 to 9 GHz

Directivity

45

52

47

Source match

36

44

34

Load match

36

45

39

Relection tracking

± 0.100

± 0.040

± 0.070

Transmission tracking

± 0.156

± 0.047

± 0.155

Transmission uncertainty (specification)

Reflection uncertainty (specification)