Atec Agilent-E5063A User Manual
Page 5
Advertising
5
Corrected system performance with type-N device connectors, 85092C electronic calibration (ECal) module
Network analyzer : E5063A
Calibration kit
: 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module
Calibration
: Full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration is not performed
Specification (dB)
Description
300 kHz to 10 MHz
10 MHz to 3 GHz
3 to 9 GHz
Directivity
45
52
45
Source match
36
44
36
Load match
36
45
38
Relection tracking
± 0.10
± 0.04
± 0.07
Transmission tracking
± 0.153
± 0.052
± 0.17
Transmission uncertainty (specification)
Reflection uncertainty (specification)
Advertising