Corrected system performance with type-N device connectors, 85092C electronic calibration (ECal) module

Network analyzer : E5063A
Calibration kit

: 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module

Calibration

: Full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration is not performed

Specification (dB)

Description

300 kHz to 10 MHz

10 MHz to 3 GHz

3 to 9 GHz

Directivity

45

52

45

Source match

36

44

36

Load match

36

45

38

Relection tracking

± 0.10

± 0.04

± 0.07

Transmission tracking

± 0.153

± 0.052

± 0.17

Transmission uncertainty (specification)

Reflection uncertainty (specification)