Deinitions

Specification (spec.):
Warranted performance. All speciications apply at 23 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the instru-
ment has been turned on. Speciications include guard bands to account for the expected statistical performance distribution,
measurement uncertainties, and changes in performance due to environmental conditions.

Typical (typ.):
Expected performance of an average unit which does not include guardbands. It is not covered by the product warranty.

General characteristics:
A general, descriptive term that does not imply a level of performance.

In this data sheet, boundary conditions are given for the speciications. For example, system dynamic range is 68 dB with the
following boundary condition.

Frequency: 300 kHz

IF bandwidth: 3 kHz

If the same boundary conditions fall under more than one category in a table, apply the best value.

This data sheet also provides technical specifications for the following calibration kits and ECal modules. For models not listed
in this data sheet, please download the free uncertainty calculator from www.agilent.com/find/na_calculator to generate the
curves for your calibration kit.

• 85032F Calibration kit
• 85033E Calibration kit
• 85052D Calibration kit
• 85092C Electronic calibration (ECal) module
• 85093C Electronic calibration (ECal) module
• N4691B Electronic calibration (ECal) module

Boundary Conditions

Calibration Kits and ECal modules