The specifications in this section apply to measurements made with the Agilent E5063A network analyzer under the following
conditions:

• No averaging applied to data
• Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature
• Response and isolation calibration performed

Corrected System Performance

System Dynamic Range

Description

Specification

Typical

System dynamic range at test port

1

(IF Bandwidth = 3 kHz)

100 kHz to 300 kHz

63 dB

300 kHz to 8.5 MHz

68 dB

8.5 to 100 MHz

91 dB

100 MHz to 4.34 GHz

92 dB

4.34 to 8.5 GHz

81 dB

8.5 to 13 GHz

75 dB

13 to 16 GHz

65 dB

16 to 18 GHz

62 dB

(IF Bandwidth = 10 Hz)

100 kHz to 300 kHz

88 dB

92 dB

300 kHz to 8.5 MHz

93 dB

97 dB

8.5 to 100 MHz

116 dB

122 dB

100 MHz to 4.34 GHz

117 dB

122 dB

4.34 to 8.5 GHz

106 dB

112 dB

8.5 to 13 GHz

100 dB

106 dB

13 to 16 GHz

90 dB

100 dB

16 to 18 GHz

87 dB

93 dB

1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effective