Atec Agilent-E5063A User Manual

Page 19

Advertising
background image

19

Analyzer environmental speciications and dimensions

Description

General characteristics

Operating environment
Temperature
Error-corrected temperature range
Humidity
Altitude
Vibration

+5 °C to +40 °C
23 °C (± 5 °C) with < 1 °C deviation from calibration temperature
20% to 80% at wet bulb temperature < +29 °C (non-condensation)
0 to 2,000 m (0 to 6561 feet)
0.21 G maximum, 5 Hz to 500 Hz

Non-operating environment
Temperature
Humidity
Altitude
Vibration

–10 °C to +60 °C
20% to 90% at wet bulb temperature < +40 °C (non-condensation)
0 to 4,572 m (0 to 15,000 feet)
0.5 G maximum, 5 to 500 Hz

Dimensions

See below

Weight (net)

11 kg

Dimensions (front view)

(in millimeters)