Atec Agilent-E5063A User Manual

Page 9

Advertising
background image

9

Corrected system performance with 3.5 mm device connector type, N4691B electronic calibration (ECal) module

Network analyzer : E5063A
Calibration kit

: N4691B (3.5 mm, 50 Ω) Electronic calibration (ECal) module

Calibration

: Full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration is not performed

Specification (dB)

Description

300 kHz to 10 MHz

10 MHz to 3 GHz

3 to 10 GHz

10 to 18 GHz

Directivity

31

46

48

46

Source match

29

41

45

42

Load match

27

42

42

39

Relection tracking

± 0.110

± 0.050

± 0.030

± 0.040

Transmission tracking

± 0.358

± 0.046

± 0.062

± 0.107

Transmission uncertainty (specification)

Reflection uncertainty (specification)

Advertising