9
Corrected system performance with 3.5 mm device connector type, N4691B electronic calibration (ECal) module
Network analyzer : E5063A
Calibration kit
: N4691B (3.5 mm, 50 Ω) Electronic calibration (ECal) module
Calibration
: Full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration is not performed
Specification (dB)
Description
300 kHz to 10 MHz
10 MHz to 3 GHz
3 to 10 GHz
10 to 18 GHz
Directivity
31
46
48
46
Source match
29
41
45
42
Load match
27
42
42
39
Relection tracking
± 0.110
± 0.050
± 0.030
± 0.040
Transmission tracking
± 0.358
± 0.046
± 0.062
± 0.107
Transmission uncertainty (specification)
Reflection uncertainty (specification)