Atec Agilent-E5063A User Manual

Page 8

Advertising
background image

8

Corrected system performance with 3.5 mm device connector type, 85052D calibration kit

Network analyzer : E5063A
Calibration kit

: 85052D (3.5 mm, 50 Ω)

Calibration

: Full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration performed

Specification (dB)

Description

100 kHz to 10 MHz

10 MHz to 3 GHz

3 to 10 GHz

10 to 18 GHz

Directivity

42

38

36

36

Source match

37

31

28

28

Load match

42

38

36

36

Relection tracking

± 0.003

± 0.004

± 0.008

± 0.008

Transmission tracking

± 0.136

± 0.100

± 0.208

± 0.328

Transmission uncertainty (specification)

Reflection uncertainty (specification)

Advertising