Atec Agilent-E5063A User Manual
Page 8
Advertising
8
Corrected system performance with 3.5 mm device connector type, 85052D calibration kit
Network analyzer : E5063A
Calibration kit
: 85052D (3.5 mm, 50 Ω)
Calibration
: Full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from
calibration temperature, isolation calibration performed
Specification (dB)
Description
100 kHz to 10 MHz
10 MHz to 3 GHz
3 to 10 GHz
10 to 18 GHz
Directivity
42
38
36
36
Source match
37
31
28
28
Load match
42
38
36
36
Relection tracking
± 0.003
± 0.004
± 0.008
± 0.008
Transmission tracking
± 0.136
± 0.100
± 0.208
± 0.328
Transmission uncertainty (specification)
Reflection uncertainty (specification)
Advertising