Measurement accuracy – Atec Rohde-Schwarz-ZVA Series User Manual

Page 10

Advertising
background image

Version 10.00, May 2012

10

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer

Measurement accuracy

This data is valid between +18 °C and +28 °C, provided the temperature has not varied by more than 1 K after calibration. Validity of
the data is conditional on the use of a suitable calibration kit. This calibration kit is used to achieve the effective system data specified
below. Frequency points, measurement bandwidth, and sweep time have to be identical for measurement and calibration (no
interpolation allowed).

Accuracy of transmission measurements
R&S

®

ZVA8

300 kHz to 1 MHz

for +15 dB to –45 dB

< 1 dB or < 6°

1 MHz to 50 MHz

for +15 dB to –30 dB

< 0.2 dB or < 2°

for –30 dB to –45 dB

< 1 dB or < 6°

50 MHz to 8 GHz

for +15 dB to +5 dB

< 0.2 dB or < 2°

for +5 dB to –55 dB

< 0.1 dB or < 1°

for –55 dB to –70 dB

< 0.2 dB or < 2°

for –70 dB to –85 dB

< 1 dB or < 6°

R&S

®

ZVA24

10 MHz to 50 MHz

for +15 dB to –30 dB

< 1 dB or < 6°

50 MHz to 400 MHz

for +15 dB to –30 dB

< 0.2 dB or < 2°

for –30 dB to –45 dB

< 1 dB or < 6°

400 MHz to 700 MHz

for +15 dB to –50 dB

< 0.2 dB or < 2°

for –50 dB to –65 dB

< 1 dB or < 6°

700 MHz to 24 GHz

for +15 dB to +5 dB

< 0.2 dB or < 2°

for +5 dB to –55 dB

< 0.1 dB or < 1°

for –55 dB to –70 dB

< 0.2 dB or < 2°

for –70 dB to –85 dB

< 1 dB or < 6°

R&S

®

ZVA40

10 MHz to 50 MHz

for +15 dB to –30 dB

< 1 dB or < 6°

50 MHz to 250 MHz

for +15 dB to –30 dB

< 0.2 dB or < 2°

for –30 dB to –45 dB

< 1 dB or < 6°

250 MHz to 700 MHz

for +15 dB to +5 dB

< 0.3 dB or < 3°

for +5 dB to –65 dB

< 0.2 dB or < 2°

for –65 dB to –80 dB

< 1 dB or < 6°

700 MHz to 2 GHz

for +15 dB to +5 dB

< 0.3 dB or < 3°

for +5 dB to –50 dB

< 0.1 dB or < 1°

for –50 dB to –65 dB

< 0.2 dB or < 2°

for –65 dB to –80 dB

< 1 dB or < 6°

2 GHz to 24 GHz

for +15 dB to +5 dB

< 0.3 dB or < 3°

for +5 dB to –55 dB

< 0.1 dB or < 1°

for –55 dB to –70 dB

< 0.2 dB or < 2°

for –70 dB to –85 dB

< 1 dB or < 6°

24 GHz to 32 GHz

for +15 dB to +5 dB

< 0.3 dB or < 3°

for +5 dB to –45 dB

< 0.2 dB or < 2°

for –45 dB to –60 dB

< 0.3 dB or < 3°

for –60 dB to –75 dB

< 1 dB or < 6°

32 GHz to 40 GHz

for +15 dB to +5 dB

< 0.4 dB or < 4°

for +5 dB to –40 dB

< 0.2 dB or < 2°

for –40 dB to –55 dB

< 0.4 dB or < 4°

for –55 dB to –70 dB

< 1 dB or < 6°

Advertising