Version 10.00, May 2012

4

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer

Specifications

Specifications apply under the following conditions: 90 minutes warm-up time at ambient temperature, specified environmental
conditions met, calibration cycle adhered to, and all internal automatic adjustments performed. "Typical values" are designated with
the abbreviation "typ.". These values are verified during the final test but are not assured by Rohde & Schwarz. "Nominal values" are
design parameters that are not assured by Rohde & Schwarz. These values are verified during product development but are not
specifically tested during production.
Unless otherwise stated, specifications apply to test ports and a nominal source power of –10 dBm.

Measurement range

Impedance

50

Test port connector

R&S

®

ZVA8

type N, female

R&S

®

ZVA24

3.5 mm, male, ruggedized

R&S

®

ZVA40

2.92 mm, male, ruggedized

R&S

®

ZVA40

2.4 mm, male, ruggedized

R&S

®

ZVA50

2.4 mm, male, ruggedized

R&S

®

ZVA67

1.85 mm, male, ruggedized

Number of test ports

2 or 4

Frequency range

R&S

®

ZVA8

300 kHz to 8 GHz

R&S

®

ZVA24

10 MHz to 24 GHz

R&S

®

ZVA40

10 MHz to 40 GHz

R&S

®

ZVA50

10 MHz to 50 GHz

R&S

®

ZVA67

10 MHz to 67 GHz

Static frequency accuracy

without optional oven quartz

8×10

–6

with optional oven quartz

1×10

–7

Frequency resolution

1 Hz

Number of measurement points

user-selectable

1 to 60001

Measurement bandwidths

1/2/5 steps

1 Hz to 1 MHz

Dynamic range of the R&S

®

ZVA8

(without optional step attenuators
and without optional direct
generator/receiver access)

from PORT 1 to PORT 2 and
from PORT 3 to PORT 4

300 kHz to 50 MHz

> 100 dB, typ. 110 dB

50 MHz to 100 MHz

> 120 dB, typ. 130 dB

100 MHz to 4 GHz

> 130 dB, typ. 140 dB

4 GHz to 7 GHz

> 125 dB, typ. 135 dB

7 GHz to 8 GHz

> 120 dB, typ. 130 dB

Dynamic range of the R&S

®

ZVA24

(without optional step attenuators
and without optional direct
generator/receiver access)

from PORT 1 to PORT 2 and
from PORT 3 to PORT 4

10 MHz to 100 MHz

> 90 dB, typ. 105 dB

100 MHz to 700 MHz

> 105 dB, typ. 120 dB

700 MHz to 2 GHz

> 125 dB, typ. 130 dB

2 GHz to 13 GHz

> 130 dB, typ. 135 dB

13 GHz to 24 GHz

> 125 dB, typ. 130 dB

Dynamic range of the R&S

®

ZVA40

(without optional step attenuators
and without optional direct
generator/receiver access)

from PORT 1 to PORT 2 and
from PORT 3 to PORT 4

10 MHz to 50 MHz

> 90 dB, typ. 100 dB

50 MHz to 500 MHz

> 105 dB, typ. 115 dB

500 MHz to 2 GHz

> 125 dB, typ. 135 dB

2 GHz to 20 GHz

> 130 dB, typ. 140 dB

20 GHz to 24 GHz

> 125 dB, typ. 135 dB

24 GHz to 32 GHz

> 120 dB, typ. 130 dB

32 GHz to 40 GHz

> 118 dB, typ. 125 dB

Dynamic range of the R&S

®

ZVA50

(without optional step attenuators
and without optional direct
generator/receiver access)

from PORT 1 to PORT 2 and
from PORT 3 to PORT 4

10 MHz to 50 MHz

> 90 dB, typ. 100 dB

50 MHz to 500 MHz

> 105 dB, typ. 115 dB

500 MHz to 2 GHz

> 125 dB, typ. 135 dB

2 GHz to 20 GHz

> 130 dB, typ. 140 dB

20 GHz to 24 GHz

> 125 dB, typ. 135 dB

24 GHz to 40 GHz

> 120 dB, typ. 130 dB

40 GHz to 45 GHz

> 115 dB, typ. 125 dB

45 GHz to 50 GHz

> 110 dB, typ. 120 dB

Advertising