Measurement speed – Atec Rohde-Schwarz-ZVA Series User Manual

Page 8

Advertising
background image

Version 10.00, May 2012

8

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer

Measurement speed

Measurement time per point

CW mode,
1 MHz measurement bandwidth

< 3.5 µs

Data transfer time

for 201 measurements points

via IEC/IEEE bus

< 2.9 ms

via VX11 over 100 Mbit/s LAN

< 1.3 ms

via RSIB over 100 Mbit/s LAN

< 0.7 ms

Switching time between channels

with no more than 2001 points

< 1 ms

Switching time between two preloaded
instrument settings

with no more than 2001 points

< 10 ms

Sweep times of the R&S

®

ZVA8, R&S

®

ZVA24, R&S

®

ZVA40 and R&S

®

ZVA50

Sweep times depend on the number of measurement points, the measurement bandwidth, and the start and stop frequencies.
They include times for retrace and internal band switching and are valid with ALC and display switched off.
Number

of

measurement

points

51 101 201 401 801 1601

R&S

®

ZVA with start frequency 5 GHz, stop frequency 5.2 GHz

For a measurement bandwidth of 100 kHz

With full one-port calibration or
with correction switched off


2.6 ms


4.0 ms


6.8 ms


12 ms


23 ms


42 ms

With TOSM calibration

3.8 ms

6.5 ms

11.6 ms

22 ms

41 ms

124 ms

For a measurement bandwidth of 1 MHz

With full one-port calibration or
with correction switched off


2.1 ms


3.0 ms


4.7 ms


8.0 ms


15 ms


26 ms

With TOSM calibration

2.8 ms

4.5 ms

7.5 ms

14 ms

26 ms

94 ms


R&S

®

ZVA with start frequency 6 GHz, stop frequency 8 GHz

For a measurement bandwidth of 100 kHz

With full one-port calibration or
with correction switched off


3.6 ms


6.4 ms


11.5 ms


19 ms


31 ms


50 ms

With TOSM calibration

4.8 ms

8.9 ms

16.3 ms

29 ms

49 ms

132 ms

For a measurement bandwidth of 1 MHz

With full one-port calibration or
with correction switched off


3.1 ms


5.4 ms


9.4 ms


14.7 ms


23 ms


35 ms

With TOSM calibration

3.8 ms

6.8 ms

12.2 ms

20.5 ms

33 ms

103 ms


R&S

®

ZVA8 with start frequency 10 MHz and stop frequency 8 GHz

R&S

®

ZVA24 with start frequency 10 MHz and stop frequency 24 GHz

R&S

®

ZVA40 with start frequency 10 MHz and stop frequency 40 GHz

R&S

®

ZVA50 with start frequency 10 MHz and stop frequency 50 GHz

For a measurement bandwidth of 100 kHz

With full one-port calibration or
with correction switched off


8.6 ms


13 ms


19.4 ms


32 ms


55 ms


92 ms

With TOSM calibration

9.9 ms

15.5 ms

25 ms

41 ms

74 ms

173 ms

For a measurement bandwidth of 1 MHz

With full one-port calibration or
with correction switched off


8.2 ms


12 ms


17.4 ms


28 ms


47 ms


75 ms

With TOSM calibration

8.8 ms

13.4 ms

20.2 ms

33 ms

57 ms

143 ms