Atec Rohde-Schwarz-ZVA Series User Manual

Page 33

Advertising
background image

Version 10.00, May 2012

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer 33

Direct generator/receiver access

These options permit direct access to the internal source output as well as to the
internal reference and measurement receiver inputs via front panel connectors.
Dynamic range with direct access utilizing these inputs is stated in the “Measurement
range” section. If all front panel jumper cables are directly connected between the
outputs and inputs, the following vector network analyzer specifications apply.

Front panel connectors

R&S

®

ZVA8 SMA,

female

R&S

®

ZVA24

2.92 mm, female

R&S

®

ZVA40

2.92 mm, female

R&S

®

ZVA50

1.85 mm, female

R&S

®

ZVA67

1.85 mm, female

Frequency range

R&S

®

ZVA8

300 kHz to 8 GHz

R&S

®

ZVA24

10 MHz to 24 GHz

R&S

®

ZVA40

10 MHz to 40 GHz

R&S

®

ZVA50

10 MHz to 50 GHz

R&S

®

ZVA67

10 MHz to 67 GHz

Dynamic range

R&S

®

ZVA8

300 kHz to 8 GHz

is reduced by 2 dB

R&S

®

ZVA24

10 MHz to 13 GHz

is reduced by 2 dB

13 GHz to 24 GHz

is reduced by 4 dB

R&S

®

ZVA40

10 MHz to 13 GHz

is reduced by 2 dB

13 GHz to 24 GHz

is reduced by 4 dB

24 GHz to 40 GHz

is reduced by 6 dB

R&S

®

ZVA50

10 MHz to 13 GHz

is reduced by 2 dB

13 GHz to 24 GHz

is reduced by 4 dB

24 GHz to 50 GHz

is reduced by 6 dB

R&S

®

ZVA67

10 MHz to 13 GHz

is reduced by 2 dB

13 GHz to 24 GHz

is reduced by 4 dB

24 GHz to 67 GHz

is reduced by 6 dB

Power range

R&S

®

ZVA8

300 kHz to 8 GHz

upper limit is reduced by 1 dB

R&S

®

ZVA24

10 MHz to 13 GHz

upper limit is reduced by 1 dB

13 GHz to 24 GHz

upper limit is reduced by 2 dB

R&S

®

ZVA40

10 MHz to 13 GHz

upper limit is reduced by 1 dB

13 GHz to 24 GHz

upper limit is reduced by 2 dB

24 GHz to 40 GHz

upper limit is reduced by 3 dB

R&S

®

ZVA50

10 MHz to 13 GHz

upper limit is reduced by 1 dB

13 GHz to 24 GHz

upper limit is reduced by 2 dB

24 GHz to 50 GHz

upper limit is reduced by 3 dB

R&S

®

ZVA67

10 MHz to 13 GHz

upper limit is reduced by 1 dB

13 GHz to 24 GHz

upper limit is reduced by 2 dB

24 GHz to 67 GHz

upper limit is reduced by 3 dB

Match R&S

®

ZVA40

10 MHz to 4 GHz

is reduced by 2 dB

Advertising