Atec Rohde-Schwarz-ZVA Series User Manual

Page 5

Advertising
background image

Version 10.00, May 2012

Rohde & Schwarz

R&S

®

ZVA Vector Network Analyzer 5

Dynamic range of the R&S

®

ZVA67

(without optional step attenuators
and without optional direct
generator/receiver access)

from PORT 1 to PORT 2 and
from PORT 3 to PORT 4

10 MHz to 50 MHz

> 70 dB, typ. 90 dB

50 MHz to 500 MHz

> 110 dB, typ. 125 dB

500 MHz to 2 GHz

> 120 dB, typ. 130 dB

2 GHz to 8 GHz

> 130 dB, typ. 140 dB

8 GHz to 12 GHz

> 125 dB, typ. 140 dB

12 GHz to 24 GHz

> 130 dB, typ. 140 dB

24 GHz to 32 GHz

> 120 dB, typ. 130 dB

32 GHz to 40 GHz

> 115 dB, typ. 125 dB

40 GHz to 50 GHz

> 115 dB, typ. 125 dB

50 GHz to 65 GHz

> 107 dB, typ. 115 dB

65 GHz to 67 GHz

> 105 dB, typ. 115 dB

67 GHz to 70 GHz

typ. 103 dB

The dynamic range is defined as the difference between the actually available maximum source power and the RMS value of the
data trace of the transmission magnitude, which is produced by noise and crosstalk with the test ports short-circuited. The
specification is valid without system error correction and at 10 Hz measurement bandwidth. The dynamic range can be increased by
using a measurement bandwidth of 1 Hz. For the R&S

®

ZVA67 at single frequencies below 2.5 GHz, the dynamic range may be

affected by spurious signals.

Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S

®

ZVA8

from PORT 1 to MEAS 2 IN

300 kHz to 10 MHz

typ. > 125 dB

10 MHz to 100 MHz

typ. > 135 dB

100 MHz to 8 GHz

typ. > 145 dB

Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S

®

ZVA24

from PORT 1 to MEAS 2 IN

10 MHz to 100 MHz

typ. > 135 dB

100 MHz to 13 GHz

typ. > 145 dB

13 GHz to 20 GHz

typ. > 140 dB

20 GHz to 24 GHz

typ. > 130 dB

Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S

®

ZVA40

from PORT 1 to MEAS 2 IN

10 MHz to 100 MHz

typ. > 140 dB

100 MHz to 20 GHz

typ. > 150 dB

20 GHz to 24 GHz

typ. > 140 dB

24 GHz to 32 GHz

typ. > 130 dB

32 GHz to 40 GHz

typ. > 120 dB

Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S

®

ZVA50

from PORT 1 to MEAS 2 IN

10 MHz to 100 MHz

typ. > 140 dB

100 MHz to 20 GHz

typ. > 150 dB

20 GHz to 24 GHz

typ. > 145 dB

24 GHz to 32 GHz

typ. > 140 dB

32 GHz to 40 GHz

typ. > 135 dB

40 GHz to 50 GHz

typ. > 130 dB

Dynamic range at optional measurement
input (direct generator/receiver access
option) of the R&S

®

ZVA67

from PORT 1 to MEAS 2 IN

10 MHz to 100 MHz

typ. > 140 dB

100 MHz to 20 GHz

typ. > 145 dB

20 GHz to 24 GHz

typ. > 145 dB

24 GHz to 32 GHz

typ. > 140 dB

32 GHz to 40 GHz

typ. > 135 dB

40 GHz to 50 GHz

typ. > 130 dB

50 GHz to 67 GHz

typ. > 125 dB

Advertising